Dryers Available
• 70 L/min
• 115 L/min
• 180 L/min
• 230 L/min
• 340 L/min
• 450 L/min
‘Optional outlet regulator & gauge assembly as above on request’ |
Applications
• AA Spectrometers
• NMR Spectrometers
• GC-ATD
• General Laboratory applications
Standard Features
> Al l P LC cont rol led
> Separate i n let and exhaust valves f i t ted
> Bench or wal l mount
> Low maintenance
> 2 years war ranty
> Qui e t operation
Options
> Add it ional i n let & out let f i l trat ion on request
> Out l e t r egulator and gauge
> Ex Hazardous area version
Benefits
> Replaces high pressure gas cyl inders
> Constant gas qua l i ty
> Compact design
> Meets the requirement s o f most TOC and FT- IR instruments
Techinical Specification
Model |
Model Inlet / Outlet
Flow l/min
(Air / N2) |
Dimensions mm (ins) |
Weight
Kg (lbs) |
|
|
Height
mm (ins) |
Width
mm (ins) |
Depth
mm (ins) |
|
CAD-070 |
70 |
500 (19.6) |
300* (11.8) |
200 (7.9) |
18 (39) |
CAD-115 |
115 |
500 (19.6) |
300* (11.8) |
200 (7.9) |
20 (44) |
CAD-180 |
180 |
600 (23.6) |
400* (15.7) |
250 (9.8) |
25 (55) |
CAD-230 |
230 |
600 (23.6) |
400* (15.7) |
250 (9.8) |
27 (59) |
CAD-340 |
340 |
600 (23.6) |
400* (15.7) |
250 (9.8) |
30 (66) |
CAD-450 |
450 |
800 (31.5) |
600* (23.6) |
300 (11.8) |
35 (77) |
Techinical Data
Ambient Temp range |
5-35°C (41-95°F) |
Max Working Pressure |
10 barg (145 psig) |
Purity |
Dewpoint: -40°C (-40°F)
Particulate: <0.1 micron
Non-methane HC <0.01ppm
Oil <0.1mg/m³ |
Electrical Supply |
220v a.c. / 1ph / 50Hz or
110v a.c. / 1ph / 50-60Hz |
Inlet / Outlet connections |
G 1/4” (BSP) Female |
|